ISO 6789:2003 parameters

The ISO 6789:2003 test is available for tool types Click wrench and Peak.

General category

  • Name: type the name of the inspection.

  • Barcode identifier: it identifies the inspection. If left blank, the identifier corresponds to the inspection name.

  • Measurement device type: STbench is the only measurement device type available for this test strategy.

  • Measurement device S/N: type the serial number of the measuring device connected to the device.

  • Tightening direction: from the drop down list select the tightening direction. The available options are:

    • Clockwise

    • Counterclockwise

  • Disable manual insert: select the check box to set the value measured by the wrench automatically as the target torque of the phase (defined by the standard ISO 6789:2003) of the maximum of the torque range of the tool.
    This parameter is shown only for Peak tools.

Phase 1 category

  • Target torque: unavailable value. The target torque is equal to a percentage value of the maximum capacity of the tool. This parameter is defined by the ISO 6789:2003 standard.

  • Torque tolerance: unavailable value. The torque tolerance is equal to percentage value defined by the ISO 6789:2003 standard.

  • Number of samples: unavailable value. The number of results to do it is equal to 5. This value is defined by the ISO 6789:2003 standard.

  • STbench channel: select the STbench channel to use to do the test. The available options are:

    • Manual selection.

    • Number of the STbench channel (from 1 to 8).

  • Transducer S/N: type the serial number of the transducer linked to the measuring device connected to the device.

Each wrench type defined by the ISO 6789:2003 standard has a different number of test points and different target torque. All the test points have the same configuration of the Phase 1 described above.
Refer to the ISO 6789:2003 standard for more information.