JJF 1610:2017

Below are the items of the JJF 1610:2017 inspection result details card:

  • Inspection: name of the inspection.

  • Route: name of the route linked to the inspection.

  • Result date: date of the inspection.

  • Measuring device: measuring device used to do the inspection.

  • Target object: name of the tool inspected.

  • Status: below are the available options, depending on the status of the inspection:

    • Partial: the inspection is closed before the batch is completed. The test can be completed later on.

    • Completed: the inspection execution is closed after the batch is completed.

    • Abandoned: the inspection execution is not completed but it is manually closed. The test cannot be completed later on.

  • Calibration points table: the test point table displays the following data:

    • Calibration points: the calibration points column shows the preload phase and the results of each calibration point of high torque and low torque rate reading.

    • U [%]: the expanded uncertainty column shows the expanded uncertainty percentage of each calibration point of high torque and low torque rate reading.

  • Statistics: below are the statistics of the execution:

    • k: coverage factor applied to the relative measurement uncertainty.

    • Notes: comment written on STbench/STpad during the test.