Data analysis for JJF 1610:2017
On the Inspection results details card, click one row of the Test point table to open the Data analysis card, related to the selected test point.
Below are the items of the Data analysis card:
Xr: mean reference value determined by the measurement device.
δ; relative tolerance.
R: repeatability.
c1: derived sensitivity coefficient.
c2: derived sensitivity coefficient.
uR: experimental standard deviation of the measurement average.
ub: standard uncertainty.
Ubrel: expanded uncertainty of the torque meter.
u: combined standard uncertainty.
U: expanded uncertainty.