ISO 6789:2003 parameters
Name (*): type the name of the inspection.
Barcode identifier (*): it identifies the inspection. If left blank, the identifier corresponds to the inspection name. When a barcode is scanned on the measuring device, the matching inspections are sent to the device.
Enable barcode regex: click the switch to enable the barcode regex feature. For more information, refer to Barcode regex.
Type: select “Tool check” from the drop down list.
Tool to inspect (*): add/select the tool to inspect.
Do one of the following:
Click Add icon (
) to add a new tool to inspect. In the Add Tool dialog box, enter the details related to the tool to inspect.
Then, click Save.
Click Search icon (
) to select a tool already configured. In the Tools workspace, select the necessary tool.
Then, click Apply.
Tool model: model of the selected tool. This item is automatically filled in after selecting the tool.
Location: location of the tool. This item is automatically filled in after selecting the tool.
Method (*): in the drop-down list, select ISO 6789:2003.
Scheduling by time: add/select/delete the scheduling.
Do one of the following:
Click Add icon (
) to add a new scheduling. In the Add Scheduling dialog box, enter the details related to the new scheduling. The Type is set to Time.
Then, click Save.
Click Search icon (
) to select a scheduling already configured. In the Scheduling workspace, select the necessary scheduling. Only the Scheduling by time are shown.
Then, click Apply.
Click Delete icon (
) to delete the selected scheduling.
ToolsNet 8
Scheduling by tool usage: add/select/delete the scheduling.
Do one of the following:
Click Add icon (
) to add a new scheduling. In the Add Scheduling dialog box, enter the details related to the new scheduling. The Type is set to Tool usage counter.
Then, click Save.
Click Search icon (
) to select a scheduling already configured. In the Scheduling workspace, select the necessary scheduling. Only the Scheduling by tool usage are shown.
Then, click Apply.
Click Delete icon (
) to delete the selected scheduling.
Scheduling by SPC failure: enabling this check, the inspection expires if an SPC linked to the Tool fails in production.
Parameters
ISO 6789 type: type of wrench defined in ISO 6789 standard.
The wrench type is defined in the configuration of the tool model.
ISO 6789 class: class of wrench defined in ISO 6789 standard.
The wrench class is defined in the configuration of the tool model.
Number of samples: type the number of tightenings to be done.
Tightening direction (*): select one of the following options:
CW: clockwise direction.
CCW: counterclockwise direction.
Measurement unit (*): select the measuring unit for the inspection.
The default measurement unit is defined in the Application settings and in the Personal settings.
Test Points
Test points: values of the test points.
They are calculated from the maximum torque of the selected tool as indicated in the ISO 6789 standard.
The test points can be given as follows:
Torque target and torque deviation
Lower limit (LSL) and upper limit (USL)
Percentage of maximum torque and torque deviation
Click either the right or the left arrow to change the view of the test points.
Operator guidance
Picture: click the icon to upload an image (max 5 MB) or a video (max 10 MB). This image is sent to STpad/STpalm/STbench.
Instructions: type text instructions to guide the operator in the inspection run.
Verification code: type the verification code that must be entered or scanned when the inspection is run. This field is case-sensitive.
Verification code description: to guide the operator, type the description of the verification code dialog box that is displayed when the inspection is run.
Assignable causes/Corrective action
Assignable cause: add/select an assignable cause. Do the following:
Click Add icon (
) to add a new assignable cause. In the Add selectable dialog box, enter the details related to the new assignable causes.
Then, click Save.
Click Search icon (
) to select an assignable cause already configured. In the Selectable workspace, select the necessary assignable cause. The displayed items are all the available assignable causes filtered from the selectable list.
Then, click Apply.
Optional: click the Optional switch to do the test without selecting the assignable cause.
Allow manual input: click the allow manual input switch to enable the possibility to add manually a value. This value is saved only for the inspection, it is not added to the assignable cause.
Corrective action: add/select an assignable cause. Do the following:
Click Add icon (
) to add a new corrective action. In the Add selectable dialog box, enter the details related to the new corrective action.
Then, click Save.
Click Search icon (
) to select a corrective action already configured. In the Selectable workspace, select the necessary corrective action. The displayed items are all the available corrective action filtered from the selectable list.
Then, click Apply.
Optional: click the Optional switch to do the test without selecting the corrective action.
Allow manual input: click the allow manual input switch to enable the possibility to add manually a value. This value is saved only for the inspection, it is not added to the corrective action.
Items marked with an asterisk (*) are required.