Measuring device parameters
Name (*): type the name of the measuring device.
Description: type a description of the measuring device.
Model (*): select a model. The available measuring devices models are:
IRC-Connect
JSB
MRTT-C
QA Station MT
Open Protocol
STa 6000
STbench
STpad
STpad Cable Adapter
STpalm
STwrench
STRwrench
The QA Station MT/Open Protocol can be created automatically by entering the QA Supervisor hostname and port (port number 4711) in the measuring device.
The Add Measuring Device dialog box changes depending on the model selected.
Serial number (*): type the serial number of the measuring device.
Identifier: type the identification code linked to the measuring device.
Location (*): select the area of the plant where using the new measuring device model. By default it is set to the is the location of the user.
Do as follows:
Click Location icon (
).
In the Location dialog box, select the area of the plant where using the new measuring device model.
Then, click Apply.Status (*): select one of the following options, depending on the status of the measuring device:
In use: the measuring device is ready to work on the production applications.
Service/Repair: the measuring device cannot be used on the assembly line before maintenance is complete. The capability of the measuring device is verified according to customer’s procedures.
Scrapped/Disabled: the measuring device is out of order. Do not use it on the assembly line.
By default, the status is set to In use
Barcode configuration (*): select one of the following options:
Online & offline: it enables the search of inspections both locally and on QA Supervisor when a barcode is scanned on the measuring device.
Only online: it enables the search of inspections on QA Supervisor when a barcode is scanned on the measuring device.
Only offline: it enables the local search of inspections when a barcode is scanned on the measuring device.
None: it disables the search of inspections by scanning a barcode on the measuring device.
This field is available only for STpad, STpalm and STbench.
Automatic clean up: click on the switch to enable/disable the deletion of the data sent by QA Supervisor before a new route is sent to the measuring device.
By enabling this switch, the results of the inspections are not deleted.
This field is available only for STpad, STpalm and STbench.Enable delete measures: click on the switch to enable/disable the deletion of measures while running an inspection.
This field is available only for STpad, STpalm and STbench.Enable custom Web links: click on the switch to enable/disable the possibility to navigate the URLs configured in the Application Settings.
This field is available only for STpad, STpalm and STbench.Enable trace processing: click on the switch to enable/disable the possibility to view and process traces on the measuring device.
This field is available only for STpad and STbench.Retrieve old results: click on the switch to enable/disable the automatic acquisition of the results stored in the measuring device when connected to QA Supervisor.
This field is available only for QA Station MT/Open Protocol.If this option is disabled, at least one measurement must be stored in the device before connecting it to QA Supervisor.
License: the license switch is displayed depending on the model of the measuring device selected. The measuring devices without network connectivity (MRTT-C, IRC-Connect, and STwrench) do not show the license switch.Click the License switch to enable the license (the switch becomes blue colored).
The counter on the right of the License switch displays the available license(s) with the total number of enabled license(s) (for example, if the counter is x/y, x is the number of the available license(s) and y is the total number of the enabled license(s)).
The License switch is disabled and the counter is 0/0, if there are no available license(s).
The License switch (and the counter) is not displayed, in case of Unlimited License.
JSB license: the JSB license switch is displayed only if the model JSB is selected.
Click the License switch to enable the license (the switch becomes blue colored).
The counter on the right of the License switch displays the available license(s) with the total number of enabled license(s) (for example, if the counter is x/y, x is the number of the available license(s) and y is the total number of the enabled license(s)).
Connectivity
Host: type the host address of the measuring device.
Port: type the port number to communicate with the measuring device.
Supplier: click the arrow to open the supplier list and select the necessary supplier.
To add a new supplier, click Add icon (
).
In the Add Supplier dialog box, enter the details related to new supplier.
Then, click Save.Purchase date: click the box and select the purchase date.
To delete a purchase date previously configured, click Delete icon (
).Calibration certificates: in the Calibration certificates box, select the necessary calibration certificate.
To add a new calibration certificate, click Add icon (
).
In the Add Calibration Certificate dialog box, enter the following details related to new calibration certificate:Calibration procedure (*): type the name of the Calibration procedure used to calibrate the measuring device.
Calibration system: calibration system of the current calibration certificate. It is the traceability information for the selected measuring device with a known accuracy, higher than the expected device under calibration and certificated according the same standard for which the calibration certification is claimed for the device under test.
Calibration system S/N: type the serial number of the system used to calibrate the measuring device.
Reference transducer: traceability information for the transducer device used to calibrate a measuring device under test, where the sensitivity and the precision is proved and certified so that its output can be used as reference values for the comparisons in the calibration procedure.
Reference transducer S/N: type the serial number of the reference transducer used.
Certificate number: type the number of the calibration certificate.
Certificate link (attachment): attachment box for any document related to the calibration certificate.
Certificate date: enter the date of the emission of the certificate.
Certification expiry date (*): enter the expiry date of the calibration certificate.
Laboratory: name of the laboratory where the calibration certificate was done.
Laboratory accreditation ID: type the accreditation identification number of the laboratory where the calibration certificate was done.
Accreditation body: type the name of the accreditation body. The Accreditation body is the independent body that assessed the laboratory for the conformity to the reference standard.
Then, click Save.
To edit a calibration certificate already configured, in the Calibration certificates box, select it.
Click Edit icon (
).
In the Edit Calibration Certificate dialog box, edit the calibration certificate details.
Then, click Save.
To delete a calibration certificate already configured, in the Calibration certificates box, select it.
Then, click Delete icon (
).Connected measuring devices: in the Connected measuring devices box, select the measuring device to be connected.
To add a new measuring device, click Add icon (
).
In the Add Measuring Device dialog box, enter the details related to new measuring device.
Then, click Save.
To select a measuring device already configured, click Search icon (
).
In the Measuring Devices workspace, select the necessary measuring device.
Then, click Apply.
To delete a measuring device already configured, in the Measuring devices box, select it.
Then, click Delete icon (
).Connected transducers: in the Connected transducers box, select the transducer to be connected with the measuring device.
To add a new transducer, click Add icon (
).
In the Add Transducer dialog box, enter the details related to new transducer.
Then, click Save.
To select a transducer already configured, click Search icon (
).
In the Transducers workspace, select the necessary transducer.
Then, click Apply.
To delete a transducer already configured, in the Connected transducers box, select it.
Then, click Delete icon (
).Notes: if necessary, type a note.
Items marked with an asterisk (*) are required.